Performance of Honeywell silicon pressure transducers
View/ Open
Date
2003Author
Vijaykumar, K.
Joseph, A.
Desai, R.G.P.
Nagvekar, S.
Prabhudesai, S.P.
Damodaran, V.
Metadata
Show full item recordAbstract
Simultaneous calibration of three Honeywell pressure transducers (PPTR) at four ambient temperatures have been carried out using a novel static calibration system developed in-house. Calibration results indicated that differing PPTRs exhibited differing zero-point offset values ranging from 2mbar to - 79mbar. The least performance was exhibited below approx. or equal to 1250mbar. Hysteresis was significant, and the mean over full-scale range varied from approx. or equal to 2mbar to 8mbar over temperature range of 10-40 degrees C. However, temperature sensitivity of zero-point offset and the slope of least-squares-fitted linear graph were negligible. Linearity was excellent, its mean over the entire pressure range being within approx. or equal to 0.006% and 0.008% over this temperature range. Zero-point offset has played a significant role in deteriorating the accuracy of the PPTR, the mean accuracy (within approx. or equal to 0.1% and -5%) having been exhibited by those having offsets of 2mbar and -79mbar respectively. The mean accuracy also displayed temperature sensitivity of approx. or equal to 1% in the range 10-20 degrees C, and negligible sensitivity beyond 20 degrees C. Use of calibration equation significantly improved the mean accuracy obtainable from the PPTR, to within -0.04% and 0.01%.